Agilent Technologies 1660AS TV Converter Box User Manual


 
Byte Position
61 40 bytes - The next 40 bytes are for Analyzer 2 Data Information. They are
organized in the same manner as Analyzer 1 above, but they occupy bytes 61
through 100.
101 26 bytes - Number of valid rows of data (starting at byte 177) for each pod.
The 26 bytes of this group are organized as follows:
Bytes 1 and 2 - Unused
Bytes 3 and 4 - Unused.
Bytes 5 and 6 - Unused.
Bytes 7 and 8 - Unused.
Bytes 9 and 10 - Unused.
Bytes 11 and 12 contain the number of valid rows of data for pod 8 of the
1660A only. Unused in the other 1660-series logic analyzers.
Bytes 13 and 14 contain the number of valid rows of data for pod 7 of the
1660A only. Unused in the other 1660-series logic analyzers
Bytes 15 and 16 contain the number of valid rows of data for pod 6 of the
1660A and 1661A only.
Bytes 17 and 18 contain the number of valid rows of data for pod 5 of the
1660A and 1661A only.
Bytes 19 and 20 contain the number of valid rows of data for pod 4 of the
1660A, 1661A, and 1662A only.
Bytes 21 and 22 contain the number of valid rows of data for pod 3 of the
1660A, 1661A, and 1662A only.
Bytes 23 and 24 contain the number of valid rows of data for pod 2 of all
models of the 1660-series logic analyzers.
Bytes 25 and 26 contain the number of valid rows of data for pod 1 of all
models of the 1660-series logic analyzers.
DATA and SETup Commands
Data Preamble Description
26–9