Agilent Technologies 1660AS TV Converter Box User Manual


 
CONNect
Command :DISPlay:CONNect {{ON|1}|{OFF|0}}
The CONNect command sets the Connect Dots mode. When ON, each
displayed sample dot will be connected to the adjacent dot by a straight line.
The waveform is easier to see in this mode. When OFF, only the sampling
points will be displayed.
Example OUTPUT XXX;":DISPLAY:CONNECT ON"
Query :DISPlay:CONNect?
The CONNect query reports if connect is on or off.
Returned Format
[:DISPlay:CONNect] {1|0}<NL>
Example OUTPUT XXX;":DISPLAY:CONNECT?"
INSert
The INSert command inserts waveforms into the current display.
Time-correlated waveforms from the logic analyzer may also be added to the
current display. The waveforms are added just below any currently displayed
signals. Only two oscilloscope waveforms can be displayed at any time.
The first parameter is optional and specifies the module from where the
waveform is to be taken. The module number is the same as the slot number
in which the master card is installed. If a module is not specified, the current
module is assumed. The second parameter is the label of the waveform that
is to be added to the current display. The label names depend on the slot in
which the acquisition cards are installed.
To insert a waveform from the oscilloscope to the oscilloscope display:
Command
:DISPlay:INSert [<module number>,]<label>
DISPlay Subsystem
CONNect
30-5