Agilent Technologies 1660AS TV Converter Box User Manual


 
Query :MACHine{1|2}:TTRigger:TERM?
<term_id>,<label_name>
The TERM query returns the specification of the term specified by term
identification and label name.
Returned Format [:MACHine{1|2}:STRAce:TERM] <term_id>,<label_name>,
<pattern><NL>
Example OUTPUT XXX;":MACHINE1:TTRIGGER:TERM? B,DATA "
TIMER
Command :MACHine{1|2}:TTRigger:TIMER{1|2} <time_value>
The TIMER command sets the time value for the specified timer. The limits
of the timer are 400 ns to 500 seconds in 16 ns to 500 µs increments. The
increment value varies with the time value of the specified timer.
<time_value> real number from 400 ns to 500 seconds in increments which vary from 16 ns
to 500 µs.
Example
OUTPUT XXX;":MACHINE1:TTRIGGER:TIMER1 100E6"
Query :MACHine{1|2}:TTRigger:TIMER{1|2}?
The TIMER query returns the current time value for the specified timer.
Returned Format
[:MACHine{1|2}:TTRigger:TIMER{1|2}] <time_value><NL>
Example OUTPUT XXX;":MACHINE1:TTRIGGER:TIMER1?"
TTRigger (TTRace) Subsystem
TIMER
22–21