Example OUTPUT XXX;":MACHINE2:COMPARE:RUNTIL EQUAL"
Query :MACHine{1|2}:COMPare:RUNTil?
The RUNTil query returns the current stop criteria for the comparison when
running in repetitive trace mode.
Returned Format [:MACHine{1|2}:COMPare:RUNTil] {OFF| LT,<value>|GT,<value>l
INRange,<value>,<value>|OUTRange,<value>,<value>|EQUal|NEQual}
<NL>
<value>
real number from −9E9 to +9E9
Example OUTPUT XXX;":MACHINE2:COMPARE:RUNTIL?"
SET
Command :MACHine{1|2}:COMPare:SET
The SET command sets every state in the reference listing to "don’t cares." If
you send the SET command by mistake you can immediately send the CLEar
command to restore the previous data. This is the only time the CLEar
command will not replace "don’t cares" with zeros.
Example OUTPUT XXX;":MACHINE2:COMPARE:SET"
COMPare Subsystem
SET
20–13