HP (Hewlett-Packard) E1418A TV Converter Box User Manual


 
HP E1418A Command Quick Reference
The following table summarizes commands for the HP E1418A.
Command Description
APPLyn :CURRent <value | MIN | MAX | DEF> Closes the output relay and outputs a current on the
specified channel,
n.
:VOLTage <
value | MIN | MAX | DEF>
Closes the output relay and outputs a voltage on the
specified channel,
n.
CALibration :CONDition? Queries the module type and calibration status. The query
returns four 16-bit unsigned integers.
:CONFigure:RESistance? Configures the module to perform calibration of the internal
calibration resistor and returns a 0 or 1 indicating the readiness
of the module.
:RESet Reloads, for module use, the calibration constants stored in
non-volatile memory.
:STORe Loads the non-volatile memory with the calibration constants
presently in use by the module.
:TEST? Performs an internal check of the calibration of each
channel and returns an integer indicating the channel
calibration condition.
:VALue:RESistance <
value>
Sets a new internal calibration resistor constant used by the
module for current output.
CALibration
n
:CONFigure:CURRent? Configures the channel for current mode calibration and
returns a 0 or 1 to indicate the channel’s readiness for
calibration.
:CONFigure:VOLTage? Configures the channel for voltage mode calibration and
returns a 0 or 1 to indicate the channel’s readiness for
calibration.
:VALue:CURRent? <
value> Enters a measured calibration value and returns a value
indicating the current state of the calibration.
:VALue:VOLTage? <
value> Enters a measured calibration value and returns a value
indicating the current state of the calibration.
DIAGnostic :CALibration:OUTPut[:MODE]
<CALBus | CHANnel>
Sets the voltage calibration path for the module to either the
CAL output terminals or each output channel.
:CALibration:OUTPut[:MODE]? Queries the voltage calibration path for the module.
:CONFigure? Returns six 16-bit words indicating the present settings of
the module.
:OUTPut:ALL[:STATe] <OFF | ON> Opens or closes all channel output relays at once.
:SOURce
n:FUNCtion:MODE
<CURRent | VOLTage>
Changes the output mode of the channel without disabling
the channel output.
:TRIGgered? Returns a 16-bit word indicating the present trigger status of
the module.
HP E1418A Command Quick Reference
Chapter 3 HP E1418A SCPI Command Reference 107