HP (Hewlett-Packard) E1418A TV Converter Box User Manual


 
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] <value | MIN | MAX | DEF>
immediately outputs the voltage set in value at channel n when the output relay is closed.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n keyword 1 to 8 or 1 to 16 1
<
value>
numeric (float 32) or
discrete
-16 V to +16 V or
MIN | MAX | DEF
0
Comments In the keyword, SOURcen, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The <value> parameter is a floating point 32-bit number in the range of -16 Volts
to +16 Volts.
MIN sets -16 Volts.
MAX sets +16 Volts.
DEF sets 0 Volts.
This command immediately programs the specified voltage on channel n. The
output relay for that channel must be closed for the voltage to appear at the output
terminals.
This command generates an error if channel n is in the current output mode.
Related Commands: SOURcen:FUNCtion:MODE VOLTage,
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]?, OUTPutn[:STATe]
SOURce
Chapter 3 HP E1418A SCPI Command Reference 91