SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] <value | MIN | MAX | DEF>
immediately outputs the voltage set in value at channel n when the output relay is closed.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n keyword 1 to 8 or 1 to 16 1
<
value>
numeric (float 32) or
discrete
-16 V to +16 V or
MIN | MAX | DEF
0
Comments • In the keyword, SOURcen, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
• The <value> parameter is a floating point 32-bit number in the range of -16 Volts
to +16 Volts.
MIN sets -16 Volts.
MAX sets +16 Volts.
DEF sets 0 Volts.
• This command immediately programs the specified voltage on channel n. The
output relay for that channel must be closed for the voltage to appear at the output
terminals.
• This command generates an error if channel n is in the current output mode.
• Related Commands: SOURcen:FUNCtion:MODE VOLTage,