HP (Hewlett-Packard) E1418A TV Converter Box User Manual


 
STATus:OPERation:CONDition?
STATus:OPERation:CONDition? returns the contents of the Operation Status
Condition Register. Reading the register does not affect its contents.
Parameters Returns: 16-bit signed integer (int16).
Comments Bit 0 of this register (decimal 1) is the Calibrating bit. This bit is set when:
the CALibration:CONFigure:RESistance?, CALibrationn:CONFigure:VOLTage?,
or
CALibrationn:CONFigure:CURRent? command has been received and,
measurements with the corresponding CALibration:VALue:RESistance,
CALibrationn:VALue:VOLTage?, or CALibrationn:VALue:CURRent? commands
are not complete.
*RST Condition: clears all conditions.
*CLS and STATus:PRESet do not affect bits in the Status Operation Condition
Register.
STATus:OPERation:ENABle
STATus:OPERation:ENABle <mask> specifies which bits of the Operational
Status Event Register are OR’d together to create the Operational Status bit (bit 7)
in the Summary Register of the Status Byte.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
mask> numeric (int16) or
non-decimal numeric
-32768 to +32767 or #H0000 to #HFFFF none
Comments Only bit 0 (decimal 0) is used or defined in the Operation Register. Bit 0 is
defined as the Calibrating bit. A 1 in this bit indicates the module is currently in
the calibration mode.
The non-decimal numeric forms are the #H, #Q, or #B formats specified by
IEEE-488.2.
*RST clears all conditions.
*CLS and STATus:PRESet do not affect bits in the Operation Status Condition
Register.
STATus
96 HP E1418A SCPI Command Reference Chapter 3