HP (Hewlett-Packard) E1418A TV Converter Box User Manual


 
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <value | MIN | MAX | DEF>
outputs the voltage set in value at channel n when the trigger conditions are satisfied
and the channel output relay is closed. Use this command to synchronize multiple
channel outputs.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n keyword 1 to 8 or 1 to 16 1
<
value>
numeric (float 32) or
discrete
-16 V to +16 V or
MIN | MAX | DEF
0
Comments In the keyword, SOURcen, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The <value> parameter is a floating point 32-bit number in the range of -16 Volts
to +16 Volts.
MIN sets -16 Volts.
MAX sets +16 Volts.
DEF sets 0 Volts.
This command sets channel n to output the voltage specified following a valid
trigger condition. Following this command, the trigger system must be enabled
using an
INITiate command. See the TRIGger subsystem discussion, page 54, for
additional information.
The channel output relay must be closed for the voltage to appear at the output
terminals.
If the module is in a wait-for-trigger state (INIT has been received) when this
command is received, an error is generated.
If the channel is in the current mode when this command is received, an error is
generated.
Related Commands: SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?,
SOURcen:FUNCtion:MODE VOLTage
SOURce
Chapter 3 HP E1418A SCPI Command Reference 93