Code Error Messages Potential Cause(s)
+3008 Invalid Calibration State An invalid sequence of calibration commands was received.
For example, a CAL
n:CONF:VOLT command followed by a
CAL
n:VAL:CURR command.
+3009 Invalid Calibration Channel
An invalid channel number,
n, was requested in the
CAL
n:CONF or CALn:VAL commands.
+3010 Calibration aborted A command, other than one of the CALibration commands, was
received during the calibration process.
+3011 Checksum error on Calibration
Memory
Calibration constants in RAM are corrupted. Recalibration is
needed.
+3012 Channel cannot be programmed to
this mode, check jumpers
An attempt to set the channel output mode when the P/J jumper
is set to Jumper selectable and the mode is set by the V/I
jumper.
+3013 Waiting for Trigger An attempt to change a channel output value or mode was
made after the trigger system in the wait-for-trigger state.
+3014 Illegal while initiated An attempt to program a channel was made when the channel
is in the wait-for-trigger state.
+3015 Non-volatile memory write failed Indicates a hardware failure in non-volatile memory.