HP (Hewlett-Packard) E1418A TV Converter Box User Manual


 
CALibrationn:CONFigure:CURRent?
CALibrationn:CONFigure:CURRent? configures the channel for current mode
calibration and returns a
0 or 1 to indicate the channel’s readiness for calibration.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n
keyword 1 to 8 or 1 to 16 1
Returns: an unsigned 16-bit integer (uint16). 1 = ready, 0 = not ready.
Comments In the keyword, CALibrationn, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
This command is used together with the CALibrationn:VALue:CURRent? <value>
query to perform current mode calibration.
The calibration is performed either at the CAL output terminals or at an individual
channel. The choice is made using the
DIAGnostic:CALibration:OUTPut[:MODE]
command.
When the DIAG:CAL:OUTP:MODE CALBus command (also the reset state)
is used, the output relay routes the calibration current to the CAL output
terminals (and opens all channel output terminals).
When the DIAG:CAL:OUTP:MODE CHANnel command is used, the output relay
routes the calibration current to the channel terminals (the CAL terminals are
not connected).
The
DIAG:CAL:OUTP:MODE command must be sent to the module before the
CALibrationn:CONFigure:CURRent? query.
This query will generate an error if the trigger system is waiting for a trigger event
when this command is received.
This query returns a 0 to indicate that the channel is not ready for calibration, and
a
1 to indicate the channel is ready for calibration.
This query is affected by the state of the program jumper for this channel on the
terminal module. To use this command:
the P/J jumper must be set to program selectable or, if set to jumper selectable,
the V/I jumper must be set to the current mode. The jumpers are described on
page 31.
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 154,
for additional information about this command.
Related Commands: CALibrationn:VALue:CURRent?,
DIAGnostic:CALibration:OUTPut[:MODE]
CALibration
Chapter 3 HP E1418A SCPI Command Reference 71