A returned value of 5 (calibration aborted) is given if either:
a. The command is received when the module is in voltage calibration or
resistance calibration mode, or
b. The calibration in process has entered more than 50 values without a
successful calibration.
If the internal calibration resistor is not calibrated, the channel is not marked as
calibrated following the completion of calibration (see the
CALibration:CONDition?
query). The new calibration constants are used, however.
When the channel current calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.
If a calibration is interrupted before completion (that is, by another command
being received), no constants will be changed.
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 154,
for additional information about this command.
Related Commands: CALibration:STORe, CALibrationn:CONFigure:CURRent?
CALibrationn:VALue:VOLTage?
CALibrationn:VALue:VOLTage? <value> enters a measured calibration value and
returns a value indicating the present state of the voltage calibration process.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n
keyword 1 to 8 or 1 to 16 1
<
value> numeric (float 32) 0 to 19.2 V none
Returns: unsigned 16-bit integer (uint16).
Comments In the keyword, CALibrationn, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
This command is used together with the CALibrationn:CONFigure:VOLTage? query
to perform voltage mode calibration. Following the
CALn:CONF:VOLT? query, this
command is used repetitively until the return value indicates that the calibration is
complete.
The <value> parameter is the measured output voltage. A minimum of
9 measurements must be taken to calibrate a voltage output channel
(a maximum of 50 measurements may be required).
CALibration