CALibration:CONFigure:RESistance?
CALibration:CONFigure:RESistance? configures the module to perform
calibration of the internal calibration resistor and returns a
0 or 1 indicating the
readiness of the module.
Parameters Returns: signed 16-bit integer (int16). 1 = ready, 0 = not ready.
Comments The internal calibration resistor should be calibrated before calibrating the
channels in current mode.
This query configures the module for a 4-wire resistance measurement at the CAL
output terminals when the calibration point is set to
CALB (following a *RST or
DIAGnostic:CALibration:OUTPut[:MODE] CALBus command).
This query may also be used to set an external calibration resistor value when
calibrating at each channel output.
This query is used together with the CALibration:VALue:RESistance command to
calibrate the current calibration resistor.
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 154,
for additional information about this command.
Related Commands: CALibration:VALue:RESistance
CALibration:RESet
CALibration:RESet reloads, for module use, the calibration constants stored in
non-volatile memory.
Parameters None.
Comments This command, with respect to the calibration constants, performs the same action
as power-on.
At power-on, or following this command, calibration constants stored in
non-volatile memory are loaded into the module where they are used during
output operations.
If the trigger system is waiting for a trigger when this command is received,
an error is generated.
This command opens all output relays.
Related Commands: CALibration:STORe
CALibration
68 HP E1418A SCPI Command Reference Chapter 3