CALibrationn:VALue:CURRent?
CALibrationn:VALue:CURRent? <value> enters a measured calibration value
and returns a value indicating the present state of the current calibration process.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n keyword 1 to 8 or 1 to 16 1
<
value>
numeric (float 32) 0.0 to 6.0 V none
Returns: unsigned 16-bit integer (uint16).
Comments In the keyword, CALibrationn, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
This command is used together with the CALibrationn:CONFigure:CURRent?
query to perform voltage mode calibration. Following the CALn:CONF:CURR?
query, this command is used repetitively until the return value indicates that the
calibration is complete.
The <value> parameter is the measured output voltage. A minimum of
9 measurements must be taken to calibrate a channel for current mode output
(a maximum of 50 measurements may be required).
This query returns one of the following values:
Return
Value
Meaning
0 Calibration is complete.
New constants are in RAM.
Resistance is calibrated.
1 More measurements are needed.
Resistance is calibrated.
2 Calibration is complete.
New constants are in RAM.
Resistance is not calibrated.
3 More Measurements are needed.
Resistance is not calibrated.
5 Calibration aborted.
CALibration