CALibration:CONDition?
CALibration:CONDition? queries the module type and calibration status. The
query returns four 16-bit unsigned integers.
Parameters Returns: four unsigned 16-bit integers (uint16).
Comments The integers returned indicate the module configuration and status at the last
calibration, not the current configuration of the module. The first and second
integers returned from this query can be compared to the first and third integers
returned from the
DIAGnostic:CONFigure? query to determine if the module
configuration has changed since the last calibration.
Integer
Returned
Meaning
1 Expansion Board ID in the form:
0 = present, 7 = none
(expansion board contains channels 9–16)
2 2nd 16-bit value indicates how the channels are configured (isolated
or non-isolated). A bit set to 0 indicates an isolated channel, a bit set
to 1 indicates a non-isolated channel. Bits 0–15 correspond to
channels 1–16, respectively.
3 3rd 16-bit value indicates the current calibration status for each
channel. A bit set to 0 indicates the channel is not calibrated for
current, a bit set to 1 indicates the channel has been calibrated for
current. Bits 0–15 correspond to channels 1–16, respectively.
4 4th 16-bit value indicates the voltage calibration status for each
channel. A bit set to 0 indicates the channel is not calibrated for
voltage, a bit set to 1 indicates the channel has been calibrated for
voltage. Bits 0–15 correspond to channels 1–16, respectively.
CALibration
Chapter 3 HP E1418A SCPI Command Reference 67