Measuring is done with a probe in intimate contact with the surface of display in the measuring part of this equip-
ment. There are 2 m(6.6ft.) and 5 m(16.4ft) long cord. There are 4 types of Measuring Probe.
Probe model Cord length Product name Probe model Cord length Product name
LED Universal Measuring ø27 Probe
2m CA-PU32
LED Flicker Measuring ø27 Probe
2m CA-P32
LED Universal Measuring ø27 Probe
5m CA-PU35
LED Flicker Measuring ø27 Probe
5m CA-P35
LED Universal Measuring ø10 Probe
2m CA-PSU32
LED Flicker Measuring ø10 Probe
2m CA-PS32
LED Universal Measuring ø10 Probe
5m CA-PSU35
LED Flicker Measuring ø10 Probe
5m CA-PS35
Either of the accessories described above is included.
1 Receptor................................................... Place this receptor against the display’s screen surface and perform
measurement.
2 Pointing ring ............................................ For zero calibration, set this ring to the “0–CAL” position to block
entry of light into the probe.
For measurement:Set the ring to the “MEAS” position to perform
measurement.
3 Ring stopper............................................. Stops the ring at two positions.
4 Screw hole ............................................... • Used to secure the probe to a jig etc.
5 Plug .......................................................... Connect this plug to the probe connector on the main unit or that
on the optional 4-Probe Expansion board (CA-B15).
6 Hood ........................................................ Used to prevent entry of ambient light and help you place the probe
at the appropriate distance (30 mm) from the display and perpen-
dicular to it.
7 Lens cap ................................................... Used to protect the receptor.
LED Universal Measuring ø27 Probe CA-PU32(2m) /CA-PU35(5m)
LED Universal Measuring ø10 Probe CA-PSU32(2m) /CA-PSU35(5m)
LED Flicker Measuring ø27 Probe CA-P32(2m) /CA-P35(5m)
LED Flicker Measuring ø10 Probe CA-PS32(2m) /CA-PS35(5m)
7 Standard Lens cap for CA-210/310 CA-H11/
Small Lens cap for CA-210/310 CA-HS11
1 Receptor
2 Pointing ring
3 Ring stopper
4 Screw hole
6 Standard Hood for CA-210/310 CA-H10/
Small Hood for CA-210/310 CA-HS10
5 Plug