Konica Minolta CA-310 CRT Television User Manual


 
54
3. Performing Matrix Calibration
Matrix calibration cannot be performed with the memory channel CH00.
(CH00 memory channel is provided for measurement that uses the Konica Minolta’s calibration standard.)
Matrix calibration must be performed for each display type (model).
Characters of displays vary with the display type (model). Because of this, measured values differ even if the
same color is measured. Thus, a different memory channel must be used for each display type (model) to
perform matrix calibration.
If matrix calibration is performed with a memory channel to which the target color has already been set, that
target color will be deleted.
If matrix calibration is performed with a memory channel which has already been white-calibrated, the correc-
tion factor of the previous white calibration will be deleted and the correction factor obtained from the matrix
calibration will be set.
If matrix calibration is performed with a memory channel for which the RGB emission characteristic for the
analyzer mode is to be set, the previous RGB emission characteristic will be deleted and the WRGB set for
matrix calibration will be set as the RGB emission characteristic.
[Operating Procedure]
When the optional 4-Probe Expansion Board CA-B15 is used
Select the probe no. to be white-calibrated. White calibration can be performed independently for each probe
connector ([P1] to [P5]) for each memory channel.
1 Press the key.
The LCD display section will switch to the menu selec-
tion screen.
2 Press the key to open the PROBE selec-
tion screen.
Each time the key is pressed, the screen will switch
in the order PROBE SYNC ID Name input
RANGE Measurement Speed Number of Digits
Calibration Standard RS232C Baud Rate
PROBE.
3 Press the key to display the probe no. you want to select.
Each time the key is pressed, the probe no. switches in the order [P1]
.
4 Press the key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automatically when the POWER switch is set to ON( | ).
Menu selection screen
PROBE selection screen
Probe no.
Press the
key
until the desired
probe no. appears.
MENU : SELECT
PUSH SPACE KEY
SELECT : PROBE
P1 35881112 A
SELECT : PROBE
P3 35881113