26
Probe connector [P1]
About Connection
1. Connecting a Measuring Probe
Before setting the POWER switch to ON ( | ), a measuring probe must be connected to the probe connector [P1] on
the instrument.
[Connecting Method]
1. Set the POWER switch to OFF (“O” position).
2. Insert the probe’s plug into the probe connector
[P1], with the probe serial no. facing down.
3. Check that the plug is inserted all the way and
connected firmly
• When disconnecting the measuring probe, set the POWER
switch to OFF(●●) first, and pull the probe by holding the
plug. Never pull the probe by its cord.
<Notes when Connecting the Probe>
● Never connect or remove the measuring probe while the POWER
switch is ON ( | ).
● When connecting/disconnecting the measuring probe, always hold
the plug and connect/disconnect it. In addition, do not pull or bend
the cord excessively or exert excessive force on it. Doing so may
result in wire breakage.
● The Measuring Luminance Range will vary according to the type
of Measuring Probe.
● When measurement is implemented, the same Measuring Probe
to be used for the User Calibration is necessary. If measurement
is carried out by connecting the different Measuring Probe, error
message E1 will be displayed.