65
Settings Section
2. Setting/changing the target color by entering values
This method can be used for memory channel CH00 only.
[Operating Procedure]
1
4
2
3
When the optional 4-Probe Expansion Board CA-B15 is used
Select the probe no. to which you want to set the target color. The target color can be set independently for each
probe connector ([P1] to [P5]) for each memory channel.
1 Press the key.
The LCD display section will switch to the menu selec-
tion screen.
2 Press the key to open the PROBE selec-
tion screen.
Each time the key is pressed, the screen will switch
in the order PROBE → SYNC → ID Name input →
RANGE → Measurement Speed → Number of Digits →
Calibration Standard→ RS232C Baud Rate → PROBE.
3 Press the key to display the probe no.
you want to select.
Each time the key is pressed, the probe no. switches in the order [P1]
…
.
4 Press the key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automatically when the POWER switch is set to ON( | ).
1. Press the MEMORY CH and keys
until the memory channel CH00 appears.
2. Press the key.
In the LCD display section, the current target color val-
ues are displayed.
<Error Messages in LCD Display Section>
…
For other error messages, refer to page 101.
● “E3” (after the
key is pressed)
• Cause : An attempt was made to set Incorrect target
color values.
Incorrect calibration values mean the following.
1 One of x, y and Lv is “0”.
2 1–x–y 0
3 Values which are beyond the instrument’s calculation capability or contradicting values.
• Action : Enter correct values and then press the
key.
Menu selection screen
PROBE selection screen
Probe no.
Press the
key
until the desired
probe no. appears.
MENU : SELECT
PUSH SPACE KEY
SELECT : PROBE
P1 35881112 A
SELECT : PROBE
P3 35881113
Memory channel
CH00 EXT Ad P1A
[ ]
CH00 x y Lv P1
3127 3293 160.0
_