Konica Minolta CA-310 CRT Television User Manual


 
14
About Measuring Probe
Setting a Measuring Probe
Two types of screws are provided to secure the measuring prove.
Tripod screw: Used to mount the prove to a tripod. The screw depth is 6 mm.
ISO screw: Used to mount the prove to a jig. An ISO screw (5 mm, depth: 6 mm) can be used.
Setting the Measuring Distance
1. Secure the display to be measured.
2. Set the pointing ring to the MEAS position.
3. Make sure that the distance from the dis-
play surface to the tip of the probe is 30
mm, and secure the probe.
Make sure that the probe is placed perpendicular to the
display surface.
<Caution>
Measurement reproducibility becomes [the way which does
not change the installation angle θ for every measurement] high in
measuring the high display of view angle dependability with the
Measuring probe.
Use of the hood (standard accessory) not only prevents entry of envi-
ronmental light but also makes it easy to place the instrument at the
specified distance and perpendicular to the object.
LED Universal Measuring ø27 Probe(CA-PU32/35), LED Flicker
Measuring ø27 Probe(CA-P32/35)
: The stated accuracy remains valid when r is in the range of 30 mm
±10 mm.
LED Universal Measuring ø10 Probe(CA-PSU32/35), LED Flicker
Measuring ø10 Probe(CA-PS32/35)
: The stated accuracy remains valid when r is in the range of 30 mm
±5 mm.
Display’s
screen
surface
(When used with the hood)
(When used without the hood)
Display
ISO screw ISO screw
Tripod screw
88 mm (116)
22 mm
98 mm (126)
108 mm (136)
*( )Measuring Ø10 Probe
Measuring Probe
θ