Konica Minolta CA-310 CRT Television User Manual


 
5050
3. Selecting the Calibration Standard (data)
This section explains how to select the instrument’s calibration standard (6500K, 9300K). Selecting the instrument’s
calibration standard will set the calibration standard for CH00 as well as for all the memory channels which have
not been user-calibrated.
[Selecting Method]
1. Press the key.
The LCD display section will switch to the
menu selection screen.
2. Press the key to open the
calibration standard selec-
tion screen.
Each time the key is pressed, the screen
will switch in the order PROBE SYNC
ID Name input RANGE Measure-
ment Speed Number of Digits Cali-
bration Standard RS232C Baud Rate PROBE.
3 . Press the key to display the calibration standard you want to select.
Each time the key is pressed, the calibration standard switches alternately between “6500K” and “9300K”.
4. Press the key to confirm the selection.
The selected calibration standard will be set for CH00 as well as for all the memory channels that have not
been user-calibrated.
* By default (factory setting), the instrument is set so that 6500K mode will be selected automatically when the POWER switch is set to
ON( | ).
*To cancel calibration standard setting, press the
key.
<Notes on Calibration Standard Setting>
The specified calibration target values will be kept even if the POWER switch is set to OFF (). The selected
calibration standard will be set for CH00 as well as for all the non-user-calibrated memory channels when the
POWER switch is set to ON( | ).
PQRS
7
TUV
8
WXYZ
MENU
ALPHA
9
GHI
4
JKL
5
MNO LOCK
6
CAL
White
1
ABC
2
DEF
-
SPACE
3
ENTER
Red
0
Green
.
Blue
MENU : SELECT
PUSH SPACE KEY
SELECT : STD-DT
6500K
SELECT : STD-DT
9300K
Press the key until the desired
calibration standard appears.
Menu selection screen
SYNC selection screen
The currently selected calibration
standard will be displayed.