Konica Minolta CA-310 CRT Television User Manual


 
55
Settings Section
[Preparation]
1. Press the MODE key to select xyLv measurement mode.
2. Press the MEMORY CH and keys until the
memory channel where you want to perform user calibra-
tion appears.
A memory channel other than CH00 must be selected.
3. Place the measuring probe against the display and take
measurement.
Set the display so that it can display four colors (RGBW) whose xyLv
values are known.
4. Press the key.
The LCD display section will switch to the user calibration input screen.
5. Enter the emission characteristic of R and calibration val-
ues (x, y, Lv).
1 Place the measuring probe against the display, which is now emitting
monochrome light of R.
Currently measured values will be displayed.
2 While the probe is placed against the display, press the HOLD key.
The measured values will be hold and the HOLD LED lights up.
3 Press the key.
The LCD display section will switch to the R calibration value input
screen.
4 Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for
white calibration (see step 7 in “Performing White Calibration” on page
52).
5 Press the key.
The LCD display section will return to the user calibration input screen,
with the “*” mark displayed on the left of “R”.
6 Press the HOLD key to resume measurement.
6. Enter the emission characteristic of G and calibration values (x, y, Lv).
1 Place the measuring probe against the display, which is now emitting monochrome light of G.
Currently measured values will be displayed.
2 While the probe is placed against the display, press the HOLD key.
The measured values will be hold and the HOLD LED lights up.
3 Press key.
The LCD display section will switch to the G calibration value input
screen.
4
Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for
white calibration (see step 7 in “Performing White Calibration” on page 53).
5 Press the key.
The LCD display section will return to the user calibration input screen,
with the “*” mark displayed on the left of “G”.
6 Press the HOLD key to resume measurement.
CH01 U-CAL
P1 W *R G B
CH01 x y Lv P1
0000 0000 .0000
_
CH01 U-CAL
P1 W *R *G B
CH01 EXT Ad P1A
[ ]
CH01 U-CAL
P1 W R G B
CH01 x y Lv P1
0000 0000 .0000
_
CH01 U-CAL
P1 W *R G B
*” mark is displayed.
*” mark is displayed.