Konica Minolta CA-310 CRT Television User Manual


 
37
Measurement Preparation
MENU : SELECT
PUSH SPACE KEY
SELECT : M-SPD
AUTO
SELECT : M-SPD
SLOW
SELECT : M-SPD
FAST
CH00 EXT Fd P1A
[ ]
Menu selection screen
Measurement speed selection screen
“F” is displayed when the [FAST] was selected.
PQRS
7
TUV
8
WXYZ
MENU
ALPHA
9
GHI
4
JKL
5
MNO LOCK
6
CAL
White
1
ABC
2
DEF
-
SPACE
3
ENTER
Red
0
Green
.
Blue
1
2
3
4
Press the key
until the desired
measurement
speed appears.
[Operating Procedure]
1. Press the key.
The LCD display section will switch to the menu selection
screen.
2. Press the key to open the measurement speed
selection screen.
Each time the key is pressed, the screen will switch in
the order PROBE SYNC ID Name input RANGE
Measurement Speed Number of Digits Calibra-
tion standard RS232C Baud Rate PROBE.
3. Press the key to display the desired measure-
ment speed.
Each time the key is pressed, the measurement speed
switches in the order [AUTO] [SLOW] [FAST]
[AUTO].
4. Press the key to confirm the selection.
* By default (factory setting), the instrument is set so that [AUTO] will be selected automatically when the POWER switch is set to
ON( | ).
* To cancel selection of measurement speed, press the
key.
<Notes when Selecting the Measurement Speed>
The selected measurement speed data will be kept even if the POWER switch is set to OFF ().
The selected measurement speed will be effective when the POWER switch is set to ON( | ).
**Flicker Mode is a function which can be used only when LED Flicker Measuring ø27 Probe(CA-P32/35) or LED Flicker Measuring ø10
Probe(CA-PS32/35) is connected.