Viscosity Meters
Electrical Connections
Another method of surge protection is to connect an MOV (Metal Oxide Varistor) (breakdown voltage
>30 V) with an NE-2 neon bulb in parallel across each wire and ground. These can be mounted in a
junction box close to the meter.
If the RS-485/Modbus output is permanently connected to a PC, an independently powered, fully
isolated RS-485 to RS-232 converter should be used. (See the Further information on RS-485 section
for details).
3.2.6 Installation in explosive areas
The meter is an explosion-proof and flameproof device. However, it is essential to observe the rules
of compliance with current standards concerning flameproof equipment:
Electronics housing caps should be tightened securely and locked in position by their locking
screws.
The electrical cable or conduit should have an appropriate explosion-proof cable gland fitted.
If any electrical conduit entry port is not used, it should be blanked off using the appropriate
explosion-proof blanking plug, with the plug entered to a depth of at least five threads.
The spigot must be locked in place.
For installations in hazardous areas:
For ATEX installations, the electrical installation must strictly adhere to the safety
information given in the ATEX safety instructions booklet shipped with this manual. See
Section 1.1 for important information.
For installations in USA and Canada, the electrical installation must strictly adhere to the
Electrical Codes and a conduit seal is required within 2” (50 mm) of the enclosure.