Emerson 7829 TV Antenna User Manual


 
54 Micro Motion 7829 Viscomaster
®
and Viscomaster Dynamic
Viscosity Meters
General Maintenance
6.2.2 Electrical check
1. Check the power supply and current consumption at the meter terminals, pins 1 and 2, having
disconnected all analog outputs. These should give 35 mA to 42 mA at 22.8 V to 25.2 V.
If the current consumption is outside this range, contact Micro Motion.
6.2.3 Performance check
When several systems are run in parallel and use the same fluid source, comparison of the line
viscosity, base density and temperature readings between installations can be a useful indicator of
possible system faults. Differences between readings, or changes from the normally observed
conditions should always be investigated to confirm that instrumentation is functioning correctly.
6.2.4 Calibration check
1. Carry out a check calibration as detailed in the Calibration Check chapter.
2. Compare the results obtained with the previous calibration figures to identify any substantial
deterioration in meter performance or any malfunction.
Note: A drop in meter performance is likely due to a build up of deposition on the tines which can be
removed by the application of a suitable solvent. See Mechanical Servicing below.
Note: Malfunctions generally could be the result of electrical/electronic faults in either the meter or
the readout equipment. Always check the readout equipment first before attention is directed to the
meter.
6.3 Fault analysis and remedial action
A fault may be categorized as either an erratic reading or a reading which is outside limits.
Electrical faults can also cause symptoms which appear to affect the readings and it is recommended
that the electrical system is checked first, before removing the meter for servicing.
6.3.1 Troubleshooting faults
Table 6-1 Faults and possible causes
Fault Possible causes Remedy
Readings fluctuate slightly,
i.e., are noisy
Analog output averaging time
not long enough
Increase the averaging time using ADView’s
Board Configuration facility (see the Using
ADView and ProLink II chapter).
Erratic readings One or more of:
Gas bubbles around tines;
cavitations; severe vibration or
electrical interference; large
amount of contaminants
Remove primary cause; e.g.:
-install air release units to release gas;
-apply back pressure to discourage formation
of bubbles;
-remove cause of vibration
Alternatively, it may be necessary to adjust
the Time Period Trap.
Readings outside limits Deposition and/or corrosion on
the tines.
Clean tines.