Emerson 7829 TV Antenna User Manual


 
46 Micro Motion 7829 Viscomaster
®
and Viscomaster Dynamic
Viscosity Meters
Using ADView and ProLink II
4.2.4 ProLink II configuration upload/download
ProLink II provides a configuration upload/download function which allows you to save configuration
sets to your PC. This allows:
Easy backup and restore of meter configuration
Easy replication of configuration sets
Micro Motion recommends that all meter configurations be downloaded to a PC as soon as the
configuration is complete.
To access the configuration upload/download function:
1. Connect ProLink II to your meter.
2. In the ProLink II software application, open the
File menu.
To save a configuration file to a PC, use the
Load from Xmtr to File option.
To restore or load a configuration file to a meter, use the
Send to Xmtr from File option.
4.2.5 ProLink II language
ProLink II can be configured for the following languages:
English
•French
•German
To configure the ProLink II language, choose
Tools > Options.
In this manual, English is used as the ProLink II language.
Connection Physical layer Protocol
RS-485 terminals or RS-485 network RS-485 Modbus