Basic Installation
TLA 700 Series Installation Manual
3
Table 1 lists the operating requirements for the logic analyzers. Refer to the
TLA 700 Series Logic Analyzer User Manual for a complete specification list.
Table 1: Mainframe operating requirements
Requirement TLA 704 Color Portable mainframe TLA 711 Color Benchtop mainframe
Source voltage 90–250 V
RMS
, 45–66 Hz, continuous range
100–132 V
RMS
, 360–440 Hz, continuous range
90–250 V
RMS
, 45–66 Hz, continuous range
100–132 V
RMS
, 360–440 Hz, continuous range
Steady state input current 6 A
RMS
, maximum 15 A
RMS
, maximum at 90 VAC
6.3 A
RMS
, maximum at 207 VAC
Maximum power consumption 600 W 1350 W line power
Temperature
(No media in floppy disk drive)
+5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
+5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
Humidity
(No media in floppy disk drive)
20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
Altitude To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
Operating Requirements