Appendix B: Upgrades
52
TLA 700 Series Installation Manual
20. Remove the module from the logic analyzer and remove the jumper from the
Flash programming pins on the rear of the module. Keep the jumper for
future updates.
21. Attach a new label with the new firmware version to the module.
22. Reinstall the module in the logic analyzer and reconnect the probes.
23. Reconnect the power cord and power on the logic analyzer.
24. Verify that the power-on diagnostics pass.
25. Run the self calibration on each module after a 30-minute warm-up.
Controller BIOS Setup
The logic analyzer has a built-in setup program where you can set some basic
input and output characteristics. In most cases, there is no need to access the
setup program. If you need to access the setup program, power on the logic
analyzer and press function key F2 on the keyboard (ALT, CTRL, F2 keys
simultaneously on the portable mainframe front panel) before the logic analyzer
boots the Windows 95 operating system.
Table 6 on page 53 lists the default settings for the BIOS setup. You may need to
refer to this information after updating the software or making other changes.
CAUTION. Do not change any of the settings if you are unfamiliar with the setup
program. Selecting an incorrect setup can cause the logic analyzer to malfunc-
tion.
If you need to change any of the settings to match those listed in Table 6, follow
the on-screen instructions.
NOTE. If you need to upgrade the Controller BIOS version, refer to the Mainte-
nance chapter of the TLA 711 Color Benchtop Controller Service Manual or the
TLA 704 Color Portable Mainframe Service Manual for the upgrade instruc-
tions.