Tektronix 070-9774-02 TV Converter Box User Manual


 
Appendix A: Merging Logic Analyzer Modules
TLA 700 Series Installation Manual
39
Merge cable
Figure 20: Dressing the Merge cable before installing the cover
CAUTION. To prevent damage to the module during the installation process,
reinstall the cover exactly as described in steps 6 through 8. If the cover is not
properly seated, the module can be damaged when you install it in a mainframe
or it will not meet EMC requirements.
6. Push forward on the cover so the tab on the front edge of the cover inserts
into the rear of the front subpanel. Make sure that the cover is fully seated
(no gaps) against the front and rear chassis flanges (see Figure 18 on
page 36).
7. While holding the cover in place, install the four screws nearest the front of
the module (two on the top and two on the bottom), to secure the cover to
the chassis.
8. Install the flat-head screws near the front of the module.
9. Slide the rear panel on the chassis and install the two rear panel screws.
10. Install the top and bottom rear screws.
11. Install the Merge cable bracket so that the guide pins point into the module.
12. Install and tighten the two screws on the Merge cable bracket.
13. Verify that you have installed and tightened all screws to 4 in lbs on the
module.