Appendix B: Upgrades
TLA 700 Series Installation Manual
51
7. Reinstall the modules in the logic analyzer.
8. Reconnect the power cord and power on the logic analyzer.
NOTE. When the TLA 700 application starts, any modules where you have
installed the Flash programming jumper will not display in the system menu.
9. Exit the TLA 700 application.
10. Click Start, point to Programs, point to Tektronix TLA 700, and click
TLA 700 Firmware Loader.
11. Select the modules that you want to update from the list of modules
displayed in the Supported list box near the top of the menu. If you are
updating more than one module at a time, note the slot location of the
module in the selection box.
12. Select Load Firmware from the Execute menu.
13. Click the proper .LOD file for the module in the C:\Program Files\ TLA 700
Firmware directory. If you are updating more than one module make sure
that you select the proper firmware file for the module in the listed slot
number.
14. Click Ok. You will be asked to confirm your action; click Yes when
prompted.
NOTE. The program will not allow you to load firmware to an incompatible
module. For example, the program will not load DSO firmware to a logic
analyzer module.
15. If you are updating another module, select the proper .LOD file for the
module.
16. Click Ok. You will be asked to confirm your action; click Yes when
prompted.
The program will load the firmware for each module one at a time. The
process takes approximately five minutes per module.
17. When the load operation is complete, select each of the modules in the
Supported list and then select Checksum from the Execute menu.
18. If each module passes the checksum test, exit the program.
19. Power off the logic analyzer and disconnect the power cord.