Emerson Process Management Rev AB Projection Television User Manual


 
Status alerts and Status Alert Severity (continued)Table 5-1:
Alert number Alert title Default severity User can reset severity
A106 Burst Mode Enabled Informational To Informational or Ignore only
A107 Power Reset Occurred Informational Yes
A113 mA Output 2 Saturated Informational To Informational or Ignore only
A114 mA Output 2 Fixed Informational To Informational or Ignore only
A115 No External Input or Polled Da-
ta
Informational Yes
A116 Temperature Overrange (API
Referral)
Informational Yes
A117 Density Overrange (API Refer-
ral)
Informational Yes
A118 Discrete Output 1 Fixed Informational To Informational or Ignore only
A120 Curve Fit Failure (Concentra-
tion)
Informational No
A121 Extrapolation Alert (Concentra-
tion)
Informational Yes
A122 Pressure Overrange (API Refer-
ral)
Informational Yes
A132 Sensor Simulation Active Informational Yes
A133 EEPROM Error (Display) Informational Yes
A136 Incorrect Display Type Informational Yes
5.5 Configure informational parameters
Display Not available
ProLink III Device Tools > Configuration > Meter Information
Field Communicator Configure > Manual Setup > Info Parameters
Overview
The informational parameters can be used to identify or describe your meter. They are not
used in process measurement and they are not required.
Procedure
Enter data as desired.
Configure device options and preferences
Configuration and Use Manual 95