Emerson Process Management Rev AB Projection Television User Manual


 
10.8 Density measurement problems
Density measurement problems and recommended actionsTable 10-3:
Problem Possible causes Recommended actions
Erratic density reading Normal process noise
Two-phase flow
Flow rate too high
Deposition on the tines
Contaminants or suspended solids in the
process fluid
Vibration in the pipeline
Erosion or corrosion
Check your process conditions.
Increase the density damping value.
Reduce the flow rate.
Check for two-phase flow. See
Section 10.22.
Ensure that line pressure or sample pres-
sure meets installation requirements.
Increase back pressure to minimize bubble
formation.
Clean the tines.
Minimize vibration in the pipeline.
Install a flow control method (bypass, flow
chamber, expander, etc.).
Perform Known Density Verification.
Inaccurate density
reading
Inaccurate temperature measurement
Incorrect calibration factors
Deposition on the tines
Boundary effect (incorrect installation)
Verify the temperature reading from the
RTD (on-board temperature sensor).
Verify the temperature reading from the
external temperature device, if applicable.
Verify or adjust the density offset and/or
the density meter factor.
Clean the tines.
Increase the flow rate.
Install a thermal insulation jacket.
Verify the calibration factors.
Perform Known Density Verification.
Change the location of the fork.
Density reading too
high
Deposition on the tines Clean the tines.
Density reading too
low
Leaks in the pipework or fittings
Two-phase flow
Check for two-phase flow. See
Section 10.22.
Cyclic inaccuracy in
density reading
Stirrer rotation rate too high
Tank flow rate too high
Fork in poor location
Stirrer creating bubbles
Solids in tank
Reduce the stirrer rotation rate.
Reduce the tank flow rate.
Move the fork to a location with a lower
flow profile.
Install a sample bypass.
Replace the device with a CDM.
Density reading from
device does not
match laboratory val-
ue
Incorrect calibration factors
Laboratory conditions do not match sam-
ple conditions
Verify the calibration factors.
Verify or adjust the density offset and/or
the density meter factor.
Ensure that the two fluids are being com-
pared at the same temperature.
Troubleshooting
174 Micro Motion
®
Fork Viscosity Meters (FVM)