Options for Discrete Output Source
Options for Discrete Output SourceTable 6-4:
Option
Label
State
Discrete output volt-
ageProLink III Field Communicator
Enhanced Event 1–5 Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
ON Site-specific
OFF 0 V
Calibration in Progress Calibration in Progress Calibration in Progress ON Site-specific
OFF 0 V
Fault (default) Fault Indicator Fault ON Site-specific
OFF 0 V
Important
This table assumes that Discrete Output Polarity is set to Active High. If Discrete Output Polarity is set to Active
Low, reverse the voltage values.
Related information
Configure an enhanced event
Fault indication with the discrete output
6.3.2 Configure Discrete Output Polarity
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > DO > CONFIG DO > DO POLAR
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Polarity
Overview
Discrete outputs have two states: ON (active) and OFF (inactive). Two different voltage
levels are used to represent these states. Discrete Output Polarity controls which voltage level
represents which state.
Procedure
Set Discrete Output Polarity as desired.
The default setting is Active High.
Integrate the meter with the control system