Emerson Process Management Rev AB Projection Television User Manual


 
Viscosity measurement problems and recommended actions (continued)Table 10-2:
Problem Possible causes Recommended actions
Viscosity reading inac-
curate
Incorrect calibration factors
Inappropriate viscosity offset or viscosity
meter factor
Flow rate too low or too high
Bypass not fully closed
Deposition or coating, especially asphal-
tenes, on the tines
Fork laminate damaged
Defective thermal insulation
Boundary effect
Verify the calibration factors. Ensure that
the viscosity ranges are appropriate for
your process fluid.
Adjust the viscosity offset or viscosity me-
ter factor.
Adjust the flow rate or the diameter of the
flow-through chamber. Refer to the instal-
lation manual for flow requirements and
best practices.
Ensure that the installation type matches
the calibration. Refer to the installation
manual.
Close the bypass.
Clean the tines.
Check for physical damage to the tines or
laminate.
Repair or replace the thermal insulation.
Check the pump delivery and service the
pump as required.
Referred viscosity
reading inaccurate
Inaccurate line viscosity measurement
Inaccurate line temperature measurement
Incorrect referred viscosity configuration
Inaccurate measurement of laboratory
sample
Ensure that line viscosity measurement is
as accurate as possible.
Ensure that the temperature value used in
referred viscosity measurement is as accu-
rate as possible.
Verify the configuration of referred viscosi-
ty measurement.
Repeat the laboratory measurement, en-
suring that reference conditions match.
Troubleshooting
Configuration and Use Manual 173