Example: Calculating the trim offset and the trim slope
Comparison 1 Laboratory value 50.00%
Meter value 49.98%
Comparison 2 Laboratory value 16.00%
Meter value 15.99%
Populate the equations:
50 =
(
A × 49.98
)
+ B
16 =
(
A × 15.99
)
+ B
Solve for A:
50.00 16.00 = 34.00
49.98 15.99 = 39.99
34 = A × 33.99
A = 1.00029
Solve for B:
50.00 =
(
1.00029 × 49.98
)
+ B
50.00 = 49.99449 + B
B = 0.00551
Concentration slope (A): 1.00029
Concentration offset (B): 0.00551