Configuration methods for referred viscosity (continued)Table 4-4:
Referred viscosity calculation
method Description
Matrix Referral Not based on ASTM D341 standards
Applicable to all process fluids
Supports measurement of two to six process fluids from one
configuration
4.6.1 Configure referred viscosity measurement, ASTM D341
Single-Curve method
Referred viscosity is line viscosity corrected to a reference temperature. In other words, this
is the viscosity that the device would report if the line temperature matched the reference
temperature. The ASTM D341 single-curve method is used only with petroleum products.
It is used when the meter will measure only one process fluid.
Configure referred viscosity measurement, ASTM D341 Single‐Curve method, using
ProLink III
Configure referred viscosity measurement, ASTM D341 Single‐Curve method, using the
Field Communicator
Configure referred viscosity measurement, ASTM D341
Single-Curve method, using ProLink III
ProLink III Device Tools > Configuration > Process Measurement > Referred Viscosity
Overview
Referred viscosity is line viscosity corrected to a reference temperature. In other words, this
is the viscosity that the device would report if the line temperature matched the reference
temperature. The ASTM D341 single-curve method is used only with petroleum products.
It is used when the meter will measure only one process fluid.
Important
Use the ASTM D341 Single-Curve method only with petroleum products.
Prerequisites
You must know the viscosity of your process fluid at two temperatures.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Referred Viscosity.
2. Set Referred Viscosity Method to ASTM D341 Single-Curve, and click Apply to refresh the
screen.