Problem Possible causes Recommended actions
Inaccurate referred
density reading
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect API table selection
Verify the line density value. If it is not ac-
curate, see Section 10.8.
Verify the line temperature value. If it is
not accurate, see Section 10.9.
Ensure that the application is configured to
use the appropriate temperature source.
Ensure that reference temperature and ref-
erence pressure, if applicable, are config-
ured correctly.
Ensure that the selected API table is appro-
priate for the process fluid.
10.11 Concentration measurement problems
Problem Possible causes Recommended actions
Significantly incorrect
concentration meas-
urement after loading
matrix
The wrong temperature or density unit
was configured when the matrix was loa-
ded
Set the temperature and density units to
the units used when the matrix was built,
then reload the matrix. For custom matri-
ces, contact Micro Motion.
Inaccurate concentra-
tion measurement
reading
Inaccurate density measurement
Incorrect matrix data
Inappropriate trim values
Verify the line density value. If it is not ac-
curate, see Section 10.8.
Ensure that the appropriate matrix is ac-
tive.
Ensure that the matrix is configured cor-
rectly.
Adjust measurement with a concentration
offset trim. See Section 9.8.
Related information
Standard matrices for the concentration measurement application
Concentration measurement matrices available by order
Load a concentration matrix using ProLink III
Troubleshooting