The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and click Change Matrix.
Note
To support matrix switching, you must select the matrix in Slot 1 or the matrix in Slot 2.
Related information
Matrix switching
4.8.3 Set up concentration measurement using the
Field Communicator
This section guides you through most of the tasks related to setting up and implementing
the concentration measurement application.
Restrictions
This section does not cover building a concentration matrix. See Micro Motion Enhanced
Density Application: Theory, Configuration, and Use for detailed information on building a
matrix.
You cannot load a concentration matrix using the Field Communicator. If you need to load a
matrix, you must use ProLink III.
1. Enable the concentration measurement application using the Field Communicator
2. Set reference temperature values for specific gravity using the Field Communicator
3. Provide temperature data for concentration measurement using the Field Communicator
4. Modify matrix names and labels using the Field Communicator
5. Modify concentration measurement operational parameters using the
Field Communicator
6. Select the active concentration matrix using the Field Communicator
Enable the concentration measurement application using
the Field Communicator
The concentration measurement application must be enabled before you can perform any
setup. If the concentration measurement application was enabled at the factory, you do
not need to enable it now.
1. Choose Overview > Device Information > Applications > Enable/Disable Applications.
2. If the API referral application is enabled, disable it.
The concentration measurement application and the API referral application cannot
be enabled simultaneously.
3. Enable the concentration measurement application.
Configure process measurement
74 Micro Motion
®
Fork Viscosity Meters (FVM)