Configure Temperature Input using the Field Communicator
Field Communicator Configure > Manual Setup > Measurements > External Inputs > Temperature
Overview
Temperature data from the on-board temperature sensor (RTD) is always available. You
can set up an external temperature device and use external temperature data if you want
to.
Tip
Use an external device only if it is more accurate than the internal RTD.
Important
Line temperature data is used in several different measurements and calculations. It is possible to
use the internal RTD temperature in some areas and an external temperature in others. The
transmitter stores the internal RTD temperature and the external temperature separately. However,
the transmitter stores only one alternate temperature value, which may be either an external
temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one
area, and digital communications in another, and configure a fixed temperature value in a third, the
fixed value will be overwritten by polling and digital communications, and polling and digital
communications will overwrite each other.
Procedure
Choose the method to be used to supply temperature data, and perform the required
setup.
Method Description Setup
Internal RTD tem-
perature data
Temperature data from the on-
board temperature sensor
(RTD) is used.
a. Choose Configure > Manual Setup > Measurements > Optional Setup
> External Inputs > Temperature.
b. Set External Temperature for Viscosity to Disable.
Configure process measurement
30 Micro Motion
®
Fork Viscosity Meters (FVM)