APPENDIX A-3: Warnings and Test Functions Teledyne API - Models T703, T703U, 703E MENU TREES (05745E
DCN6572)
A-18
APPENDIX A-3: Warnings and Test Functions
Table A-2: Warning Messages
NAME
1
MESSAGE TEXT DESCRIPTION
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the CPU was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED Configuration storage was reset to factory configuration or
erased.
WPHOTOLTEMP
PHOTO LAMP TEMP
WARNING
Photometer lamp temperature outside of warning limits specified
by PHOTO_LAMP variable.
WO3GENTEMP
O3 GEN LAMP TEMP
WARNING
O
3
generator lamp temperature outside of warning limits
specified by O3_GEN_LAMP variable.
WPHOTOREF
PHOTO REFERENCE
WARNING
Photometer reference reading less than 2500 mV or greater than
4999 mV.
WLAMPSTABIL
PHOTO LAMP STABILITY
WARNING
Photometer lamp reference step changes occur more than 25%
of the time.
WO3GENREF
O3 GEN REFERENCE
WARNING
O
3
reference detector drops below 50 mV during reference
feedback O
3
generator control.
WO3PUMP
O3 PUMP WARNING O3 pump failed to turn on within timeout period specified by
O3_PUMP_TIMEOUT variable.
WBOXTEMP
BOX TEMP WARNING Chassis temperature outside of warning limits specified by
BOX_SET variable.
WREARBOARD
REAR BOARD NOT DET
Rear board was not detected during power up.
WRELAYBOARD
RELAY BOARD WARN
Firmware is unable to communicate with the relay board.
WLAMPDRIVER
LAMP DRIVER WARN Firmware is unable to communicate with either the O
3
generator
or photometer lamp I
2
C driver chip.
WFRONTPANEL
FRONT PANEL WARN
Firmware is unable to communicate with the front panel.
WANALOGCAL
ANALOG CAL WARNING
The A/D or at least one D/A channel has not been calibrated.
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.