Teledyne API T703/T703U Calibrator Operation Manual Getting Started
41
Table 3-6: Possible Warning Messages at Start-Up
MESSAGE MEANING
ANALOG CAL WARNING
The calibrator’s A/D converter or at least one D/A
channel has not been calibrated.
CONFIG INITIALIZED
Stored Configuration information has been reset to the
factory settings or has been erased.
DATA INITIALIZED
The calibrator’s data storage was erased.
LAMP DRIVER WARN
The firmware is unable to communicate with either the
O
3
generator or photometer lamp I
2
C driver chips.
O3 GEN LAMP TEMP WARNING
The O
3
generator lamp temperature is outside of
allowable limits.
O3 GEN REFERENCE WARNING
The O
3
generator’s reference detector has dropped
below the minimum allowable limit.
O3 PUMP WARNING
The pump associated with the O
3
photometer has failed
to turn on.
PHOTO LAMP TEMP WARNING
The photometer lamp temperature is outside of
allowable limits.
PHOTO REFERENCE WARNING
The photometer reference reading is outside of
allowable limits.
REAR BOARD NOT DET
The calibrator’s motherboard was not detected during
power up.
RELAY BOARD WARN
The firmware is unable to communicate with the
calibrator’s relay board.
SYSTEM RESET
The calibrator has been turned off and on or the CPU
was reset.
07223C DCN6572