Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
96
4.12.5.2. Editing or Deleting a LEVEL
To edit or delete an existing LEVEL, press:
Levels are displayed according to the following Format:
LEVEL ID: Any number between 0 and 99. This will be the number used to select
the level when activating / deactivating it, or when editing or deleting it.
Gas Conc: The concentration setting, in ppb, for the O
3
generator to produce.
Status Block Setting: This will be displayed as two pairs of 1-digit numbers.
The First pair corresponds to Status Block 1.
The Second pair corresponds to Status Block 2.
n each case:
The left digit will be a number between 1 and 4 representing the binary setting
of bits 1 through 4 and;
The right digit will be a number between 1 and 4 representing the binary setting
of bits 5 through 8.
Figure 4-5: LEADS Level Display Format
07223C DCN6572