Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
50
4.3. STANDBY MODE
When the T703 Photometric Calibrator is in standby mode, it is at rest.
Some functions under the SETUP MORE DIAG submenu, those which conflict
with accurate creation of calibration gas mixtures (e.g. ANALOG OUTPUT STEP
TEST) automatically place the calibrator into STANDBY mode when activated
NOTE
The T703/T703U calibrator should always be placed in STANDBY mode when not needed to produce
calibration gas.
This can be done manually by pressing the STBY button that appears when the calibrator’s display is
showing the top-level menu (see Figure 4-1).
When programming a calibration sequences the STANDBY step should always be inserted at the end of
the sequence.