Teledyne T703U VCR User Manual


 
Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
48
4.1. TEST FUNCTIONS
A variety of TEST FUNCTIONS are available for viewing at the front panel whenever
the calibrator is at the MAIN MENU. These measurements provide information about
the present operating status of the calibrator and are useful during troubleshooting (see
Section 8). Table 6-1 lists the available TEST functions.
To view these TEST func
tions, press <TST or TST> from the main menu and the Test
Functions will scroll at the top center of the display.
Table 4-1: Test Functions Defined
TEST MEASUREMENT DESCRIPTION
O3GEN FRAC=X.XX (T703U only) For low range mode to produce low levels of ozone during
a GPT calibration where O
3
output is <500 PPB LPM
(supported only in GEN-AUTO mode)
O3 PRES=XX.X PSIG
O
3
generator regulator pressure
ACT=GENERATE XX PPB O3 Actual concentration being generated, computed from
real-time inputs
TARG=GENERATE XX PPB O3
Target concentration to generate
OUT FLW=X.XXX LPM
Output flow rate (computed from regulator pressure)
REG PRESS=XX.X PSIG
Regulator pressure
BOX TMP=XX.X ºC
Internal chassis temperature
OGEN REF=XXXX.X MV
O
3
generator reference detector reading
O3GENDRV=XXXX.X MV
O
3
generator lamp drive output
O3LAMPTMP=XX.X ºC
O
3
generator lamp temperature
PH MEAS=XXXX.X MV
Photometer detector measure reading
PH REF=XXXX.X MV
Photometer detector reference reading
PH FLW=X.XXX LPM
Photometer sample flow rate
PH LTEMP=XX.X ºC
Photometer lamp temperature
PH SPRES=XX.X IN-HG-A
Photometer sample pressure
PH STEMP=XX.X ºC
Photometer sample temperature
PH SLOPE=X.XXX Photometer slope computed during zero/span bench
calibration
PH OFFST=X.X PPB Photometer offset computed during zero/span bench
calibration
PHOTO STABIL=X.X PPB Photometer concentration stability (standard deviation of
25 bench concentration samples taken 10 seconds apart)
TEST=XXXX.X MV Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable
TIME=HH:MM:SS
Current instrument time of day clock. (24 hour format)
07223C DCN6572