4.9.2.3. Manual Calibration of the TEST Channel Configured for Voltage Ranges
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
NOTE:
The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL feature
is turned off for the channel being adjusted (see Section 4.9.2.1)
Calibration is performed with a voltmeter connected across the output terminals (See
Figure 4-2) and by changing the actual output signal level using the front panel menu
buttons in
100, 10 or 1 count increments.
V
+DC Gnd
Figure 4-3: Setup for Calibrating the TEST CHANNEL
Table 4-10: Voltage Tolerances for the TEST CHANNEL Calibration
FULL
SCALE
ZERO
TOLERANCE
SPAN VOLTAGE
SPAN
TOLERANCE
MINIMUM
ADJUSTMENT
(1 count)
0.1 VDC ±0.0005V 90 mV ±0.001V 0.02 mV
1 VDC ±0.001V 900 mV ±0.001V 0.24 mV
5 VDC ±0.002V 4500 mV ±0.003V 1.22 mV
10 VDC ±0.004V 4500 mV ±0.006V 2.44 mV