5q f-21
Batch finish output
off
0: Off at next start
(Turning off until next start)
1: Off at over or unstable.
(Turning off when "out of range" or
"unstable condition")
2: Off at zero band.
(Turning off at zero band)
0 to 2 0
Category address
:
[Function] - [Function setting] - [Sequence] - [Timer]
Category
address
symbol
Name Descriptions
Range
and
choices
Default
5q f-31
Batch monitoring
timer
Set the upper limit of the time of
sequence.
If the sequence is not finnished,
sequence error SQ.ERR4 is displayed.
If 0 is set, the timer does not work.
0 to
65535 s
0 s
5q f-32
Batch start delay
timer
Set the delay time between start
operation and the flow start
0.0 to
65535 s
0.0 s
5q f-33
Full flow
comparison
interrupt timer
0.0 to
25.5 s
0.0 s
5q f-34
Medium flow
comparison
interrupt timer
0.0 to
25.5 s
0.0 s
5q f-35
Dribble flow
comparison
interrupt timer
The timer to prevent weighing error due
to vibration of open/close operation.
While the timer is not up, the
comparison does not work.
0.0 to
25.5 s
0.0 s
5q f-36
Drible flow auto-
Free Fall override
time
Set the time until the dribble flow is can
stant.
Use to prevent compensation error of
automatic free fall compensation or
real time free fall compensation.
0.0 to
25.5 s
3.0 s
5q f-37
Eval delay timer
Set the time between closing dribble
flow and comparing the result. Refer to
Eval conditions [5q f-13]
0.1to
25.5 s
0.1 s
5q f-38
Discharge start
delay timer
Set the time between operating
discharge start and openning the
discahrge gate
0.0 to
25.5 s
0.0 s