Page 24 AD-4402
5.4.
5.4.5.4.
5.4. Calibration Error
Calibration ErrorCalibration Error
Calibration Error
Error Code Treatment and Situation
CERR1
Resolution (Weighing capacity / minimum division) is exceeds the limitation.
Increase minimum division or decrease weighing capacity.
CERR2
The initial load (no load output) is larger than 2mV/V.
Confirm the loadcell cable.
CERR3
Negative loadcell output value. Check wiring.
Confirm the loadcell cable.
CERR4
Mass value exceeds the weighing capacity.
Use a mass within the weighing capacity. (Decrease mass value)
CERR5
Mass value is too light for the calibration.
Increase mass value.
CERR6
The loadcell output to be equivalent to minimum division is too small.
Use more rough minimum division.
CERR7
The polarity of loadcell output is inversed.
Confirm the loadcell cable.
CERR8
The mass value of the weighing capacity exceeds 3.2 mV/V.
Confirm the mass and weighing capacity.
CERR9
Gravity acceleration is out of range.
Correct the value within the range of 9.770 ~ 9.835 m/s
2
.
CERR10
Zero output of loadcell unit is out of range.
Trim the zero output within 0.0 ~ 2.0 mV/V.
CERR11
The loadcell output to be equivalent to minimum division is out of range.
Trim the output within 0.0 ~ 3.2 mV/V.