A&D PD4000243 Universal Remote User Manual


 
AD-4402 Page 39
Making alarm when the sequence is time over.
[5q f-31] Maximum weighing time between start and batch finish can be set.
Error code [SQ.ERR 4] is displayed, when an error occurs.
[Function] - [Function setting] - [Sequence] - [Timer] - [Batch monitoring
timer]
Removing "stable" from comparison condition.
[5q f-13]
[Function] - [Function setting] - [Sequence] - [Control] - [Eval condition]
Changing the timing of comparison.
[5q f-37]
[Function] - [Function setting] - [Sequence] - [Timer] - [Eval delay timer]
Changing accuracy of comparison.
[5q f-48] The time to average weighing value at batch finish can be set. The
timing of batch finish delays for the time.
[Function] - [Function setting] - [Sequence] - [Timer] - [Average Eval time]
Changing the pulse width of weighing finish output.
[5q f-43] If zero is set to this, the output leaves until next start signal.
[Function] - [Function setting] - [Sequence] - [Timer] - [Batch finish output
on]
Mixing it at weighing finish.
[5q f-14]
[Function] - [Function setting] - [Sequence] - [Control] - [Batch finish action]
Discharging it at weighing finish.
[5q f-15]
[Function] - [Function setting] - [Sequence] - [Control] - [Discharge finish
action]
Using customer programmed control for hi signal, go signal and low signal.
[5q f- 5]
[Function] - [Function setting] - [Sequence] - [Basic] - [Comparison]