Emerson Process Management MLT 1 TV Converter Box User Manual


 
2 - 6
AK
90003752(3) [AK-Commands] 11/03
2.2. Read commands
CODE
Function Page
AAEG Kn Spangas deviation 2 - 39
AALI Kn Mm Deviations of the last linearization checks with spangas 2 - 40
AANG Kn Zerogas deviation 2 - 41
AEMB Kn Selected range (R1, R2, R3, R4) 2 - 45
AFDA Kn CODE Zero- /spangas time (function length of calibration) 2 - 46
AFDA Kn CODE Purge time 2 - 46
AGID Kn Device identification 2 - 47
AGRW Kn m Limit 2 - 48
AIKG Kn Concentration integral average; all (in ppm) 2 - 49
AIKO Kn Concentration integral average (in ppm) 2 - 51
AKAK Kn Mm Calibration gas concentration (in ppm) 2 - 53
AKAL Kn Stored calibration corrections 2 - 54
AKFG Kn Configuration of the system 2 - 57
AKON Kn Concentration (current value in ppm) 2 - 58
ALCH Kn Mm Deviations of the last linearization checks 2 - 61
ALIK Kn Calculation of linearization curve 2 - 62
ALIN Kn Mm Linearization values in the device (X/Y = Setpoint-/raw value) 2 - 63
ALKO Kn Mm Polynomial coefficents of the linearization curve 2 - 64
ALST Kn Linearization steps 2 - 65
AM90 Kn Actual response time (t90 time) 2 - 66
AMBA Kn Mm Begin of range (in ppm) 2 - 67
AMBE Kn Mm End of range (in ppm) 2 - 68
AMBU Kn Switch levels for autoranging 2 - 69
AMDR Kn Manual adjusted pressure 2 - 70
AQEF Kn Cross interference check result (in ppm) (only for CO analyzers) 2 - 71
ASOL Kn Setpoint value with limits 2 - 72
ASTA Kn General status of the system 2 - 73
ASTF Kn Internal error status 2 - 74
ASTZ Kn Device status 2 - 76
ASYZ Kn System time (year, month, day, hour, min., sec.) 2 - 78
AT9O Kn T90 time (response time) 2 - 79
ATEM Kn m Temperature 2 - 80
ATOL Kn Mm Stability tolerances 2 - 81
AVEZ Kn Delay and synchronization time 2 - 83
AZEI Kn CODE Times (for procedures) 2 - 84
Optional:
ABST Kn Counter of operating hours 2 - 42
ADRU Kn m Pressure (for service) 2 - 43
ADUF Kn m Flow (for service) 2 - 44
AKEN Kn Device tag 2 - 56
AKOW Kn Mm Correction (zerogas calibration and gradient) 2 - 60
AUKA Kn Uncorrected analog value 2 - 82
Kn:
Channel n
Mm:
Range m
Response
The response to the read command will contain the CODE of the read command,
the error status byte (0-9) and the data.