SATK – Control command "Automatic calibration"
Starting this command the analyzers in a system or the single analyzer will start a
calibration procedure to determine the correction values. The required calibration gases
and the pumps will be switched on automatically. Then, the calibration procedure will start
automatically. Such a procedure has to run until the end. It may not be canceled or
interrupted by other functions. Otherwise it is not sure that the correction values will be
valid to calculate the exhaust values of analysis. Exceptions are the commands "Reset" or
"Stand-by". After the procedure will be over the system, the analyzers in a system or the
single analyzer will have to change to the operation mode "Stand-by".
Control command
SATK K0
Function associated to the whole system unit
Code
SATK K1 Kn
Function associated to channel n
Function associated to channel 1
Code
SATK Kn M3
Function associated to channel n and range 3
Code
Response
SATK 0
Error status
Code