Emerson Process Management MLT 1 TV Converter Box User Manual


 
II) V24/RS232/485 Interface - Single Analyzers and Systems
90003752(1) [AK-Commands] 10/98
AK
2 - 41
AANG – Read command "Deviation to zerogas"
To this read command the analyzers in a system or the single analyzer will send to the
TBCC the following data for the called channel (device):
The determined and stored signal of the last zerogas measurement with its range.
The deviation from the setpoint value of the linearized curve in ppm and percent,
referred to the end of range value.
Zerogas measurement: Signal after the end of the functions "Automatic calibration" or
"Zerogas", stability controlled or time controlled depending on the setup in "EFDA".
Read command
AANG K0
AANG Kn
Read of the whole system unit
Code
Response
AANG 0 M1 XXX YYY ZZ ... Mn XXX YYY ZZ
Deviation [%]
Deviation [ppm]
Signal [ppm]
Range n
Deviation [%]
Deviation [ppm]
Signal [ppm]
Range 1
Error status
Code
The values will get the same format for the read of single channels.